1420 1 Jun 67
TTY0 doesnn't turn main shaft is locked
SRR
1 June 67
The P-clock, Sprach Dac, Eyr (III), and reference supply, are now on their own AC buss + 728 power units.
In a very short time all A-D-MPX inputs will go to the Moby-path kludge in the top of Bay (2).
To make it as flexible as poss. I have made each group of (9) chammels split into 3 subgroups of from 1-9 channels.
So long as these subgroups don't share the same channel nothing can go wrong!!
It is expected that the users cable be pluged in to one or more subgroup slots, and they be patched in to the proper channel on the wireing side of Bay Two.
Also there is (3) dual output reference supplys that appear on the wireing side of that bay, so that, they to can be patched to the user cable.
Power Supply | # | 1 | 2 | 3 | L→R on [back?] |
---|---|---|---|---|---|
All in Bay(2) | +10 | B30L | B31L +8-10 | B32L | |
-10 | B30N | B31N 0--10 | B32N | ||
GRND | B30P | B31P GND | B32P |
47
June 2-67
Repaired 3 model 33 teletypes Set# 52515 double t[r?]iping adjusted linkage Set# 53385 Replaced distributor Hub and belt Set# 53384 Replaced Belt and 1-18[11?]20 Gear.
Jesse Gonzales
June 2-67 2400
TOOK 1 (100 FT) ROLL OF 16 MM TRI-X IN COMPUTER ROOM OF FOLLOWING:
130 DISPLAY - SPACEWAR + TEXT MODE/ IT WAS HARD TO METER BUT MADE FOUR 10 SEC EXP. AT DIFF. F & SPD.
TV MONITOR - TRIPED 3 DIFF SPD TO SEE EFFECTS
COMPUTER ROOM AS A WHOLE --
CLOSE UP OF CHESS BOARD -- TO CHECK DEPTH OF FEILD AS WELL AS EXPOSURE
+ A FEW OTHER SHOTS IN ROOM TO USE REMAINDER OF FILM
THE BULK OF FILM WAS SHOT AT 24 FPS WITH THE EXCEPTION OF TV, + DISPLAY SO THE RUN TIME WILL BE 2 MIN 44 SEC I THINK, ENOUGH TO CHECK THE RESULTS
Randy
P.S. THE FIRST ROLL WAS EXPOSED AT ASA 200 WHICH IS NORM FOR MOVIE FILM AFTER WE SEE RESULTS I MAY PUSH IT TO 300 --> 400 ASA ALSO HAVE SOME ANSCOCROME/200 ASA COLOR TO TRY BUT WANT TO SEE FIRST ROLL OF B/W FILM FIRST
48
0120 - 7-JUN-67
The plotter operated on FORTRAN output with plotter-[get?] punch switch on the punch; the operation was on the order of 1/2 normal speed
R Nedy
0925 8 June 67
Above is not a but, it is a feature!
SRR
TTY3's left leg needs fixing MH[H?]
8 June 10:00
DTA3 gives random DATA ERRORS on [types?] which seem OK on other drives
[Doyle?]
49
6/8 67
Results of 100' roll of TRI-X in computer room
Scene & sub lens | F Stop | 5 PD | Distance to film plane | Results |
---|---|---|---|---|
5W 130 Dis 50 mm TV lens | 1.4 | 24 FPS | 4'9" | overexposeure |
[ditto marks] | 2 | [ditto marks] | [ditto marks] | better |
6/8 67
QI connected to I/O bus Device 734 System still works? Device not checked out, so do not use.
Lou
1930 8 Jun 67
Cleaned MAG TAPE tested it for 5 min. It still works.
SRR
300 9 June 67
LPT has trouble hanging onto forms (top sprockets)
C. Farrell
50
6-9. Films: TRI-X 200 ASA B/w 16mm Reversa
First four scenes were of 130 display
Shocking space war; filmed under following parameters: Lens 50 mm 1.4 from TV camera Distance 5'-9" to film plane Dark room -- First two exposures 1.4 @ 24 FPS 2 @ 24 FPS Next [two exposures] 1.4 @ 18 frames per second 2 @ [18 frames per second]
1.4 @ 24 was clearly overexposed 2 @ 24 was better but ships were still slightly overex.
Note @ 24 FPS there was a definet flicker of display (Display refresh rate is 30 P/s)
1.4 @ 18 was poor 2 @ 18 was better but should be set 1 stop down for best results.
Note @ 18 FPS the flicker was not as noticeable as above
Sujested exposure for above film (ASA 200) F 2.8 @ 18 FPS
If flicker is a factor slow down to 16 FPS and do ness. F conversion